MCO-230AIC
MCO-230AIC
• Incu saFe interior
• SafeCell UV system (option)
• Accurate temperature and CO2 level control
• Double stackable
Additional information
External dimension | 770 x 730 x 905 (mm) / 30.3 x 28.7 x 35.6 (inch) |
---|---|
Internal dimension | 643 x 523 x 700 (mm) /25.3 x 20.6 x 27.6 (inch) |
Internal volume | 230 liters (8.1 cu.Ft.) |
Data acquisition | Automatic log function of temperature, CO2, Door opening/closing, Alarm and CSV file output |
Exterior cabinet & door | Galvanized steel with baked-on finish |
Interior & shelves | Copper-enriched stainless steel |
Outer door | Galvanized steel with baked-on finishReversible heated |
Inner door | Tempered glass |
Shelves | Copper-enriched stainless steel 4 supplies as standard (Maximum 10), Exterior dimensions: 620 (W) x 450 (D) x 12 (H) mm, maximum load 7kg/shelf, Adjustable |
Acces ports | Diameter 30mm port with non-VOC silicone stoppers (1 on back side) |
Insulation | Extruded polystyrene foam |
Heating method | Direct Heat & Air Jacket System |
Air flow | Gentle vertical airflow, continuous with inner door closed |
Air filter | 0.3 μm, Efficiency: 99.97 % (for CO2) |
UV lamp | – |
Water level type | – |
Communication | Remote alarm contacts standard. Optional 4-20mA connection. Optional with RS-232C/ RS-485/LAN data ports |
CO2 gas connection | 4mm to 6mm inner diameter tubing |
CO2 gas pressure | 0.03 MPa (G) (0.3Kgf/cm2G, 4.3psiG) from two stage CO2 regulator |
Net weight | 90kg (198 lbs) |
Temp. control range | 5°C above ambient to 50 °C*1 (Ambient temperature: 5°C-35°C) |
Temp. Uniformity | ±0.25°C (23°C ambient, setting: 37°C, CO2: 5%, no load)*2 |
CO2 control range & deviation | 0% to 20%,±0.15% (23°C ambient, setting 37°C, 5% CO2, no load) |
Interior humidity | 95% ±5%R.H. at 37°C by natural evaporation with humidifying pan |
Max. power consumption | Max. 440W |
Max heat discharge | 1,250kJ/h |
CO2 sensor platform | Ceramic based, single beam infrared sensor, with dual wavelength measurement for continuous auto-zero calibration |
CO2 sampling patent pending | No moving parts; airflow passess over in/out ports to sustain continuous sampling |
Leveling feet | 4 |
CO2 calibration | – |