PHCbi
MCO-230AIC

PHCbi
MCO-230AIC


• Incu saFe interior
• SafeCell UV system (option)
• Accurate temperature and CO2 level control
• Double stackable

catalogue2

Compare
Categories: ,
000

Additional information

External dimension

770 x 730 x 905 (mm) / 30.3 x 28.7 x 35.6 (inch)

Internal dimension

643 x 523 x 700 (mm) /25.3 x 20.6 x 27.6 (inch)

Internal volume

230 liters (8.1 cu.Ft.)

Data acquisition

Automatic log function of temperature, CO2, Door opening/closing, Alarm and CSV file output

Exterior cabinet & door

Galvanized steel with baked-on finish

Interior & shelves

Copper-enriched stainless steel

Outer door

Galvanized steel with baked-on finishReversible heated

Inner door

Tempered glass

Shelves

Copper-enriched stainless steel 4 supplies as standard (Maximum 10), Exterior dimensions: 620 (W) x 450 (D) x 12 (H) mm, maximum load 7kg/shelf, Adjustable

Acces ports

Diameter 30mm port with non-VOC silicone stoppers (1 on back side)

Insulation

Extruded polystyrene foam

Heating method

Direct Heat & Air Jacket System

Air flow

Gentle vertical airflow, continuous with inner door closed

Air filter

0.3 μm, Efficiency: 99.97 % (for CO2)

UV lamp

Water level type

Communication

Remote alarm contacts standard. Optional 4-20mA connection. Optional with RS-232C/ RS-485/LAN data ports

CO2 gas connection

4mm to 6mm inner diameter tubing

CO2 gas pressure

0.03 MPa (G) (0.3Kgf/cm2G, 4.3psiG) from two stage CO2 regulator

Net weight

90kg (198 lbs)

Temp. control range

5°C above ambient to 50 °C*1 (Ambient temperature: 5°C-35°C)

Temp. Uniformity

±0.25°C (23°C ambient, setting: 37°C, CO2: 5%, no load)*2

CO2 control range & deviation

0% to 20%,±0.15% (23°C ambient, setting 37°C, 5% CO2, no load)

Interior humidity

95% ±5%R.H. at 37°C by natural evaporation with humidifying pan

Max. power consumption

Max. 440W

Max heat discharge

1,250kJ/h

CO2 sensor platform

Ceramic based, single beam infrared sensor, with dual wavelength measurement for continuous auto-zero calibration

CO2 sampling patent pending

No moving parts; airflow passess over in/out ports to sustain continuous sampling

Leveling feet

4

CO2 calibration

Slot Thailand