MCO-170AIC
• Incu saFe interior
• SafeCell UV system (option)
• Accurate temperature and CO2 level control
• Double stackable
Specification
EXTERNAL DIMENSION | : | 620 x 710 x 900 (mm) / 24.4 x 28.0 x 35.4 (inch) | |||
INTERIOR DIMENSION | : | 490 x 523 x 665 (mm) / 19.3 x 20.6 x 26.2 (inch) | |||
INTERNAL VOLUME | : | 165 liters (5.8 cu.Ft.) | |||
DATA ACQUISITION | : | Automatic log function of temperature, CO2, Door opening/closing, Alarm and CSV file output | |||
EXTERIOR CABINET & DOOR | : | Galvanized steel with baked-on finish | |||
INTERIOR & SHELVES | : | Copper-enriched stainless steel | |||
OUTER DOOR | : | Galvanized steel with baked-on finishReversible heated | |||
INNER DOOR | : | Tempered glass | |||
SHELVES | : | Copper-enriched stainless steel 4 supplies as standard (Maximum 10), Exterior dimensions: 475 (W) x 450 (D) x 12 (H) mm, maximum load 7kg/shelf, Adjustable | |||
ACCES PORT | : | Diameter 30mm port with non-VOC silicone stoppers (1 on back side) | |||
INSULATION | : | Rigid polyurethane foamed | |||
HEATING METHOD | : | – | |||
AIR FLOW | : | Gentle vertical airflow, continuous with inner door closed | |||
AIR FILTER | : | 0.3 μm, Efficiency: 99.97 % (for CO2) | |||
UV LAMP | : | – | |||
WATER LEVEL TYPE | : | – | |||
COMMUNICATION | : | Remote alarm contacts standard. Optional 4-20mA connection. Optional with RS-232C/ RS-485/LAN data ports | |||
CO2 GAS CONNECTION | : | 4mm to 6mm inner diameter tubing | |||
CO2 GAS PRESSURE | : | 0.03 MPa (G) (0.3Kgf/cm2G, 4.3psiG) from two stage CO2 regulator | |||
NET WEIGHT | : | 80kg (176 lbs) | |||
TEMP. CONTROL RANGE | : | 5°C above ambient to 50 °C*1 (Ambient temperature: 5°C-35°C) | |||
TEMP. UNIFORMITY | : | ±0.25°C (23°C ambient, setting: 37°C, CO2: 5%, no load)*2 | |||
CO2 CONTROL RANGE & DEVIATION | : | 0% to 20%,±0.15% (23°C ambient, setting 37°C, 5% CO2, no load) | |||
INTERIOR HUMIDITY | : | 95% ±5%R.H. at 37°C by natural evaporation with humidifying pan | |||
MAX. POWER CONSUMPTION | : | Max. 379W | |||
MAX HEAT DISCHARGE | : | 1,070kJ/h | |||
CO2 SENSOR PLATFORM | : | Ceramic based, single beam infrared sensor, with dual wavelength measurement for continuous auto-zero calibration | |||
CO2 SAMPLING PATENT PENDING | : | No moving parts; airflow passess over in/out ports to sustain continuous sampling | |||
LEVELING FEET | : | 4 | |||
CO2 CALIBRATION | : | – |